Abstract

ABSTRACTThe inductively coupled plasma (ICP)-based method provides the opportunity to determine phosphorus (P) along with other major, secondary, micro and trace elements in plant materials. This study was conducted to compare and evaluate the relative efficacy of the inductively coupled plasma-optical emission spectrometry (ICP-OES) method with that of the colorimetric method using Skalar autoanalyzer, for determining P in 428 grain samples of eight diverse crops. The results on grain P analysis by the two methods, for individual as well as for all crop samples combined, showed that they were highly positively correlated (r varied from 0.84 to 0.98, p < 0.0001 for the eight crops, and R2 for all crop grain samples was 0.9201, p < 0.0001). Moreover, the precision by the ICP method was similar to that determined by the Skalar method. Our results demonstrate that the ICP-OES method can be conveniently used for determining P along with other plant nutrient elements in grain samples of diverse crops.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.