Abstract

ABSTRACT Background To evaluate the performance of BACTEC FX, BacT/ALERT 3D, and VIRTUO systems using simulated blood culture (BC). Methods Two experimental designs based on ‘with’ or ‘without’ added trough antibiotic concentrations in bottles were implemented. Results For the experiment A, A shorter time to detection (TTD) was observed for most of organisms (17/22) in VIRTUO system. VIRTUO system was also faster than 3D and FX systems no matter in aerobic and anaerobic bottles. The anaerobic bottles had faster detection than aerobic bottles in 3D system (13.68 h vs 15.36 h, P < 0.001) and VIRTUO system (10.30 h vs 12.46 h, P = 0.001) but not in FX system (P = 0.38). When antibiotics were present, the bacterial recovery rate (RR) of FX, 3D and VIRTUO systems were 64.10% (50/78), 58.97% (46/78) and 43.59% (34/78), respectively (P = 0.027). the bacterial RR of various bottles were as follows: BPA vs. FA vs. SA [84.44%(38/45) vs. 55.56%(25/45) vs. 42.22(19/45), P < 0.001]; BFN vs. FN vs. SN [36.36%(12/33) vs. 63.64%(21/33) vs.45.45%(15/33), P = 0.078]. Conclusions The VIRTUO system allowed faster growth detection for most of organisms compared with FX and 3D systems. When antibiotics were present, the bottles containing antibiotic-binding agent showed better bacterial RR, especially in BACTEC Plus Aerobic/F bottles.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.