Abstract
A comparative dosimetric study has been done on three low Z TL phosphors: LiF TLD-100, LiF (Mg,Cu) and LiF (Mg,Cu,P). The main TL (dosimetric) peak occurs at almost the same temperature (~205 oC) in all these phosphors. The relative TL sensitivities (peak height method) to 60Co gamma rays of LiF TLD-100, LiF (Mg,Cu) and LiF (Mg,Cu,P) were found to be 1, 3.3 and 25.4 respectively. Using our TLD reader system, the respective minimum measurable gamma ray doses were estimated to be 51, 34 and 4.6 µGy. The effect of different thermal treatments on the TL sensitivity was also studied.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.