Abstract

To investigate the effect of energy (kVp) and filters (no filter, half Bowtie, and full Bowtie) on the dose response curves of the Gafchromic XRQA2 film and nanoDot optical stimulated luminescence dosimeters (OSLDs) in CBCT dose fields. To measure surface and internal doses received during x-ray volume imager (XVI) (Version R4.5) and on board imager (OBI) (Version 1.5) CBCT imaging protocols using these two types of dosimeters. Gafchromic XRQA2 film and nanoDot OSLD dose response curves were generated at different kV imaging settings used by XVI (software version R4.5) and OBI (software version 1.5) CBCT systems. The settings for the XVI system were: 100 kVp∕F0 (no filter), 120 kVp∕F0, and 120 kVp∕F1 (Bowtie filter), and for the OBI system were: 100 kVp∕full fan, 125 kVp∕full fan, and 125 kVp∕half fan. XRQA2 film was calibrated in air to air kerma levels between 0 and 11 cGy and scanned using reflection scanning mode with the Epson Expression 10000 XL flat-bed document scanner. NanoDot OSLDs were calibrated on phantom to surface dose levels between 0 and 14 cGy and read using the inLight(TM) MicroStar reader. Both dosimeters were used to measure in field surface and internal doses in a male Alderson Rando Phantom. Dose response curves of XRQA2 film and nanoDot OSLDs at different XVI and OBI CBCT settings were reported. For XVI system, the surface dose ranged between 0.02 cGy in head region during fast head and neck scan and 4.99 cGy in the chest region during symmetry scan. On the other hand, the internal dose ranged between 0.02 cGy in the head region during fast head and neck scan and 3.17 cGy in the chest region during chest M20 scan. The average (internal and external) dose ranged between 0.05 cGy in the head region during fast head and neck scan and 2.41 cGy in the chest region during chest M20 scan. For OBI system, the surface dose ranged between 0.19 cGy in head region during head scan and 4.55 cGy in the pelvis region during spot light scan. However, the internal dose ranged between 0.47 cGy in the head region during head scan and 5.55 cGy in the pelvis region during spot light scan. The average (internal and external) dose ranged between 0.45 cGy in the head region during head scan and 3.59 cGy in the pelvis region during spot light scan. Both Gafchromic XRQA2 film and nanoDot OSLDs gave close estimation of dose (within uncertainties) in many cases. Though, discrepancies of up to 20%-30% were observed in some cases. Dose response curves of Gafchromic XRQA2 film and nanoDot OSLDs indicated that the dose responses of these two dosimeters were different even at the same photon energy when different filters were used. Uncertainty levels of both dosimetry systems were below 6% at doses above 1 cGy. Both dosimetry systems gave almost similar estimation of doses (within uncertainties) in many cases, with exceptions of some cases when the discrepancy was around 20%-30%. New versions of the CBCT systems (investigated in this study) resulted in lower imaging doses compared with doses reported on earlier versions in previous studies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.