Abstract

The electrochemical behaviour of alloys (Au50-Ag25-Pd25 and Ni88.6-Cr11.4) was studied in Fusayama's artificial saliva at pH 6.5 and 37 °C by using open circuit potential, electro­chemical impedance spectroscopy, and potentiodynamic polarization measure­ments. Electrochemical impedance spectroscopy results were simulated with an equivalent electrical circuit. After immersion in artificial saliva, surface characterization of samples was done using scanning electron microscopy connected with energy-dispersive spectro­scopy. All obtained results revealed that Au50-Ag25-Pd25 alloy is much more resistive than Ni88.6-Cr11.4 and can be recommended for the effective treatment of patients with dental prosthetics that have metal frameworks. s.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.