Abstract

ObjectivesTo evaluate whether the routine coagulation tests can be performed using platelet depleted plasma (PDP, residual platelet count <40000/μL) to achieve maximum efficiency of the automated workflow and compare results of these tests performed with platelet poor plasma (PPP residual platelet count <10,000/μL) prepared manually ‘offline’. Design and MethodsThe PDP was obtained first following ‘on line’ centrifugation at 4150 RPM (3000g) for 7 min. The routine coagulation tests, Prothrombin Time (PT), Activated Partial Thromboplastin Clotting Time (aPTT), D-dimer (DD), Antithrombin III (AT3) and Fibrinogen (FBG) were performed. The PPP was obtained from an aliquot of PDP samples with additional ‘manual off line’ centrifugation at 7700 RPM (3314g) for 3 min (total 10 min, online + offline) and the same tests were performed. The statistical analysis was carried out using EP Evaluator v11 to compare results from both methods. ResultsThe results from both PPP and PDP samples demonstrated strong correlation. For example, PT (R = 0.9989; N = 55, and of Bias −0.12 (−0.67%), aPTT(R = 0.9957; N = 60, Bias 0.26 (0.58%)), AT3(R = 0.9800; N = 49, Bias −2.0 (−2.2%)), FBG (R = 0.9956; N = 57, Bias −1.9 (−0.5%)) and DD (R = 0.9981; N = 38, Bias 0.005 (0.373%)) with insignificant bias. ConclusionsThe utilization of the Roche cobas® 8100 automated ‘online’ centrifugation helps achieve optimal workflow efficiency without impacting analytical performance of the PT, aPTT, DD, AT3 and FBG assays. The use of PDP can be superior method to PPP for routine coagulation tests.