Abstract

We hypothesized that the loop material and size could affect the results of the culture when compared to the calibrated pipette. A total of 484 urine samples were included in the study, and each sample was plated by using different loop types and the calibrated pipette. The bacterial counts per milliliter were calculated and compared, with a focus on the important cutoff values of 10³ and 10⁴ CFU/ml for further identification. When considering the 10³ CFU/ml as cutoff value, 1 µl and 10 µl plastic loops gave the highest sensitivity (86.8 %), whereas the 10 µl metal loop had the lowest sensitivity (64.2 %). For the 10⁴ CFU/ml cutoff value, 1 µl plastic loop inoculation demonstrated the highest sensitivity (75.9 %), while the 10 µl metal loop provided the lowest sensitivity (26.5 %). These results suggest that the single use plastic loops are functional, sensitive, useful especially for critical sample.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.