Abstract
ObjectiveTo compare two portable lactate devices, Lactate Scout and StatStrip Xpress, to a laboratory reference device by assessing the concordance of cord blood lactate values at birth. MethodsWe conducted a single-center prospective validation study in a level 3 maternity hospital. For all cord samples taken at birth, we analyzed lactates on the reference device (GEM4000®) and on two portable devices Lactate Scout and StatStrip Xpress. We compared the lactate values from each of the two handheld devices to the reference device. ResultsA total of 101 samples were collected. Each was analyzed by all three machines. The average lactate value obtained by the different machines was 3.7mmol/L. The mean difference between the Lactate Scout and GEM4000® apparatus was 0.0mmol/L±0.8mmol/L. Between these two devices at lactate threshold of 4.8mmol/L, the negative predictive value was 96,2% (76/79) and the area under the curve was 0.98 (95% CI 0.96–1]. The mean difference between StatStrip Xpress and GEM4000® was 0.1mmol/L±0.7mmol/L. Between these two devices at the lactate threshold of 4.8mmol/L, the negative predictive value was 97,4% (76/78) and the area under the curve was 0.95 (95% CI 0.86–1). ConclusionThere is a good correlation between the lactates obtained by the reference device and the two portable devices Lactate Scout and StatStrip Xpress.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.