Abstract

Higher blood pressure levels are associated with vascular dysfunction as early as childhood. Here, we aim to compare two blood pressure devices for use in childhood populations within a school setting to screen for elevated blood pressure in children. Systolic and diastolic blood pressure, mean arteriolar pressure (MAP) as well as heart rate (HR) were measured with the validated Oscillomate 9002 and Mindray VS-900 in 82 randomly selected children between 10 and 12 years taking part in the EXAMIN YOUTH study in 2020. Measurements were performed after 5 min at rest in a sitting position in the school environment. Both devices showed strong correlations for systolic (r = 0.62, t[80] = 7, p < 0.001) and diastolic blood pressure (r = 0.73, t[80] = 9.52, p < 0.001), MAP (r = 0.75, t[80] = 10.22, p < 0.001) and HR (r = 0.89, t[79] = 17.77, p < 0.001). There was no evidence for a statistically significant difference of both devices for systolic (110 ± 9 mmHg vs. 111 ± 9 mmHg, p = 0.574), diastolic blood pressure (67 ± 9 mmHg vs. 66 ± 9 mmHg, p = 0.301) as well as MAP (81 ± 8 mmHg vs. 80 ± 8 mmHg, p = 0.400) and HR (83 ± 12 BPM vs. 83 ± 11 BPM, p = 0.994). Both devices, the older Oscillomate 9002 and the current Mindray VS-900 showed good agreement for the measurement of blood pressure in school children. It therefore appears to be feasible to measure childhood blood pressure with either device or replace one device with another in prospective long-term studies or screening programmes as long as both are validated for use in children.