We designed, fabricated and tested a novel compact fluorescence analysis system for quantification of uric acid (UA) in clinical samples at the point-of-care. To perform an analysis, diluted saliva, urine or blood samples are simply placed in a disposable thin-film sample holder using a dropper. A new enzyme immobilization technique was developed to retain within the sample holder two enzymes and a molecule, which transforms into a fluorescer in amounts depending on the UA concentration. The small instrument (7.5 cm × 5 cm × 5 cm) into which the sample holder is placed for analysis contains an LED, a narrow-band filter and an amplified photodiode. The analysis time is 30 s, and the dynamic range of the system is 4–400 μM of UA. The calibration curve for transparent saliva and urine was made using solutions of UA. The calibration curve for opaque blood was obtained with spiked samples of blood. The three different types of clinical samples were collected from three subjects and simply diluted before their measurements. Analysis with our instrument yielded UA concentrations within the expected concentration ranges. Development of instruments based on the current laboratory prototype is expected to result in products for clinical trials and point-of-care.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.