Abstract

Compact nuclear magnetic resonance (NMR) spectroscopy combined with chemometric tools opens new perspectives for NMR use. This work compares the potential of 43, 60 and 400 MHz NMR spectroscopy for quality control of edible oils. Partial least squares regression (PLSR) and support vector regression (SVR) models built on the three NMR devices had equivalent performances for fatty acids and iodine value, and the models built with the low field spectra were equivalent to the high field. Moreover, performances for calibration indicated that most of the models built with medium/or high-resolution fields presented reproducibility values lower than the minimum accepted by the American Oil Chemists' Society (AOCS). Compared to classical methods, this new approach allows the application of medium resolution devices as a sample screening tool in analytical laboratories since it allows the spectrum obtention in a few seconds, without the need for sample preparation or the use of deuterated solvents.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.