We demonstrate a flow-meter sensor using a bare laser diode (LD) as a self-mixing interferometer and no additional lens nor optical component. The diode is placed directly on the edge of the duct, exploiting the natural laser divergence. The flow measurement is based on the Doppler shift induced by the scattering particles inside the fluid and read by the self-mixing effect inside the LD. The proposed approach has been modeled by a numerical simulator in order to find the best signal processing algorithm. A set of measurements on blood proves good performances as a flow meter, while simulations confirm the system robustness to mechanical misalignments and changes in liquid properties.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.