Abstract
We have developed a compact hard X-ray imaging system composed of a cadmium telluride double-sided strip detector (CdTe-DSD) and a coded mask. We investigate the imaging performance using two different coded masks with different sizes and patterns. In our system, a CdTe-DSD of pitch 250μm is used in conjunction with a coded mask is placed 70-100 mm above the detector to form a compact imaging system. We obtained an angular resolution of up to 11.8 arc min, as measured from gamma-ray lines of point-like radioactive isotope sources. This is consistent with that expected from the geometry. The energy resolution is 1.7 keV (FWHM) at 60 keV and the energy range of imaging is from 5 keV to 122 keV. These results agree very well with Monte Carlo simulations of the detector.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.