Abstract

A grazing incidence reflectometer operating in the extreme ultraviolet (EUV) spectral range around 13.5 nm is presented which is making use of a compact xenon pinch plasma light source. The apparatus allows for measuring the absolute reflectivity of a sample for grazing incidence angle in the range from typically 5° to 35° by comparing the EUV diode signal for the reflected light and a reference diode with an accuracy of better than 2%. Design criteria for proper matching of diode apertures and distances with respect to the spatially extended plasma source are presented. The absolute accuracy has been checked by investigating a ruthenium sample with low roughness, which has a reflectivity in the EUV close to the theoretical limit. Comparison to measurements at the EUV-reflectometer of the Physikalisch Technische Bundesanstalt in Berlin at a synchrotron source confirm the absolute accuracy of better than 2% for the reflectivity for the angle interval of interest.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.