Abstract

Automation of irradiated bismuth target dissolution and astatine recovery from nitric acid media has been achieved at Texas A&M University. This process can be controlled remotely; it does not require any chemical treatment or evaporation of dissolution solution prior to the separation of above-mentioned elements on an extraction chromatography column, making the final product ready within 20 min. The accompanying radio-impurities have been identified and successfully separated from astatine, leading to greater than 99% radiopurity of the desired fractions. The system consists of only one pump and a set of LabVIEW controlled valves, and allows the connection of up to 10 different columns, providing the opportunity to prepare a variety of air dry columns, which can be different column geometries, containing astatine ready for further chemistry or shipment to desired facilities.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.