Abstract

A new compact accelerator mass spectrometry (AMS) system has been installed in the Kaminoyama research institute at Yamagata University. The AMS system is based on a 0.5 MV Pelletron accelerator developed by National Electrostatics Corp. The performance of the system was investigated using C series samples (C1‐C8), standard samples (HOxII), and reagent graphite without any chemical treatment. The precision of 14C measurements for the standard samples is typically higher than 0.3%. The ratio of 14C to 12C is less than 6×10−16 for the reagent graphite. In this paper, we present the performance of the new compact AMS system, as well as of the fully automated 20‐reactor graphite lines equipped at the research institute.