Abstract

Certified Reference Material 470 (CRM 470) demonstrates commutability with both the manufacturer's calibrator and with dilutions of serum pools in the Dade Behring N High Sensitivity assay for C-reactive protein (CRP). Both regression and back calibration show similar nonlinearity for all materials, largely due to the method of calibration curve fitting used in this assay. Significant differences in values among the currently available commercial assays can be largely overcome by using appropriate calibration curve fitting and the recommended value transfer protocol, which includes a minimum of two assay runs on each of at least 3 separate days, with weight correction of all reconstitutions and dilutions. An initial weight-corrected dilution should be made each day because of the relatively high level of CRP in CRM 470. In our opinion, the degree of nonlinearity, imprecision, and differences in values in currently available assays renders the use of fixed clinical decision cut-points questionable for high-sensitivity CRP. An alternative approach is suggested.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.