Abstract

Graphite pellet vaporization was evaluated as a method for solid sample introduction for ICP-MS. A detection limit of 0.1 ppb of Cd was obtained indicating that the technique might be very useful for screening. The accuracy was poor using the technique of external standards and four reference materials, suggesting that the technique of standard additions might be required if the technique was to be used for anything other than semi-quantitative measurements.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.