Abstract

AbstractThis experiment was conducted to identify common mode currents that flow on the shield of a coaxial cable and to investigate techniques that reduce their effects at high fields of ≥3 Tesla. It will be shown that there are two different methods of common mode interference for two different coil types. First, is the transceive case, where common mode cable shield currents flow as a result of unbalanced loop voltages that couple to the cable shield. In the second case, unwanted common mode signals are induced from the field of the transmission coil onto the cable shield of the receive‐only coil. Two important factors relating to the investigation of common mode signals are danger to the patient from an RF burn and the reduction of the signal‐to‐noise ratio (SNR) because of extra noise being introduced into the MRI environment. It will be shown that through the use of cable shield traps and balun matching circuits, common mode signals can be reduced significantly, increasing patient safety and SNR. © 2003 Wiley Periodicals, Inc. Concepts Magn Reson Part B (Magn Reson Engineering) 19B: 1–8, 2003.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.