The stentless aortic bioprosthesis has been used because of its excellent hemodynamics and few valve-related complications. We report a case of redo aortic root replacement for severe aortic regurgitation and dilatation of the Valsalva sinus 7years after the implantation of a Prima Plus aortic root bioprosthesis (Edwards LifeScience, Irvine, CA, USA) using a full root technique. Intraoperative findings showed the complete detachment of the commissure between the left and non-coronary cusps, and Valsalva sinus dilatation of the porcine aortic root bioprosthesis. Redo aortic root replacement with a 23-mm porcine bioprosthesis and 28-mm straight graft was performed. There were no findings of intimal tear, suture dehiscence, degeneration, and perforation of the bioprosthesis. Such complications associated with the Edwards Prima Plus aortic root bioprosthesis were rarely reported. Commissural detachment of a porcine stentless aortic bioprosthesis can occur; thus, careful follow-up involving echocardiography and computed tomography is necessary.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.