Abstract
The Intrabeam is a mobile intraoperative radiotherapy system that can be used to treat selected breast cancer patients with a highly localised dose of radiation (50 kV). The commissioning project involved verifying the key dosimetric parameters of the Intrabeam including (A) absorbed dose, (B) percentage depth dose and (C) dose distribution. The absolute dose determined for all applicators at a clinically relevant depth (5 mm in water) satisfied a distance to agreement criteria of 1 mm (i.e. the dose fall off is approximately 12% per mm at 5 mm depth, and agreement with the calibration files was within 12%). Excellent agreement (
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.