Abstract

We report progress for the ERDA time-of-flight beamline at the Centre of Micro-Analysis of Materials in Madrid. This beamline experienced a long shut down and has recently gone through a re-commissioning process. The time stations, based on a twin mirror design, demonstrate a time resolution of 215 ps for 6 MeV He2+. Using a standard particle detector for the energy signal we find a mass resolution of 1.4 atomic mass units for 27Al and 28Si with 40 MeV Cl4+. Detection efficiency needs no corrections for elements heavier than 7Li. Analysis of a borosilicate reference showed good agreement with the known composition. Based on these results, the facility is considered operational and open for users.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call