Abstract

MedAustron is a synchrotron-based particle therapy centre located in Wiener Neustadt, Austria. It features three irradiation rooms for particle therapy, where proton beams with energies up to 252.7MeV and carbon ions of up to 402.8MeV/u are available for cancer treatment. In addition to the treatment rooms, MedAustron features a unique beamline exclusively for non-clinical research (NCR). This research beamline is also commissioned for proton energies up to 800MeV, while available carbon ion energies correspond to the ones available in the clinical treatment rooms.Based on the requirements for particle therapy, all irradiation rooms offer particle rates of up to 109particles/s for protons and 107particles/s for carbon ions. However, for research purposes, lower particle fluxes are required and were therefore commissioned for the NCR beamline. Three particle flux settings with particle rates ranging from ≈ 2.4 × 103particles/s to ≈ 5.2 × 106particles/s were established for seven proton energies below 252.7MeV. In addition to the particle rate, the spot sizes and beam energies were measured for these settings. Furthermore, three low flux settings for 800MeV protons with particle rates ranging from ≈ 2 × 103particles/s to ≈ 1.3 × 106particles/s were commissioned. Since the commissioned low flux settings are in a regime well below the limits of the available standard beam diagnostics, setting up the beam under these new operational conditions entirely relied on the use of external detectors. Furthermore, a beam position measurement based alignment without using the standard beam profile monitors was performed for 800MeV protons.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call