Abstract

The purpose of this study was to evaluate the surfaces of commercially pure titanium (cp Ti) implants modified by laser beam (LS), without and with hydroxyapatite deposition by the biomimetic method (HAB), without (HAB) and with thermal treatment (HABT), and compare them with implants with surfaces modified by acid treatment (AS) and with machined surfaces (MS), employing topographical and biomechanics analysis. Forty-five rabbits received 75 implants. After 30, 60, and 90days, the implants were removed by reverse torque and the surfaces were topographically analyzed. At 30days, statistically significant difference (P<0.05) was observed among all the surfaces and the MS, between HAB/HABT and AS and between HAB and LS. At 60days, the reverse torque of LS, HAB, HABT, and AS differed significantly from MS. At 90days, difference was observed between HAB and MS. The microtopographic analysis revealed statistical difference between the roughness of LS, HAB, and HABT when compared with AS and MS. It was concluded that the implants LS, HAB, and HABT presented physicochemical and topographical properties superior to those of AS and MS and favored the osseointegration process in the shorter periods. In addition, HAB showed the best results when compared with other surfaces.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call