Abstract

We evaluated a commercially available, double-indicator-dilution densitometric system for the estimation of pulmonary extravascular water volume in oleic acid-induced pulmonary edema. Indocyanine green and heavy water were used as the nondiffusible and diffusible tracers, respectively. Pulmonary extravascular water volume, measured with this system, was 67% of the gravimetric value (r = 0.91), which was consistent with values obtained from the radioisotope methods. The measured volume was not influenced by changes in cardiac index over a range of 1 to 4 L.min.m2. This system is less invasive than the thermal-dye technique and has potential for repeated clinical measurements of pulmonary extravascular lung water and cardiac output.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call