Abstract
Uncontrolled hemorrhage secondary to unstable pelvic fractures is a preventable cause of prehospital death in the military and civilian sectors. Because the mortality rate associated with unstable pelvic ring injuries exceeds 50%, the use of external compression devices for associated hemorrhage control is paramount. During mass casualty incidents and in austere settings, the need for multiple external compression devices may arise. In assessing the efficacy of these devices, the magnitude of applied force has been offered as a surrogate measure of pubic symphysis diastasis reduction and subsequent hemostasis. This study offers a sensor-circuit assessment of applied force for a convenience sample of pelvic compression devices. The SAM® (structural aluminum malleable) Pelvic Sling II (SAM Medical) and improvised compression devices, including a SAM Splint tightened by a Combat Application Tourniquet® (C-A-T; North American Rescue) and a SAM® Splint tightened by a cravat, as well as two joined cravats and a standard-issue military belt, were assessed in male and female subjects. As hypothesized, compressive forces applied to the pelvis did not vary significantly based on device operator, subject sex, and subject body fat percentage. The use of the military belt as an improvised method to obtain pelvic stabilization is not advised.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of special operations medicine : a peer reviewed journal for SOF medical professionals
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.