Abstract

In the paper, "Investigation of histology region in dielectric measurements of heterogeneous tissues," by Porter and O'Halloran, the authors utilize a flexible phantom in a layered material dielectric property analysis to quantify the effective sensing volume of a coaxial dielectric probe. Ostensibly, this test has been used by others to characterize the region for which percent variation in the material composition in front of the probe corresponds to percent variation in the computed effective dielectric properties. By employing a compressible material, the authors fail to isolate features that are attributable solely to the probe, itself, and inadvertently incorporate confounding characteristics associated with the compressible nature of the material. The net effect is to exaggerate the probe's sensing volume which undermines conclusions drawn from the subsequent tissue dielectric property studies.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.