Abstract

Hochmeister et al. (1) performed validation tests for the Hexagon OBTI kit in 1999 and found a sensitivity of 1:100,000 when diluted in sterile water, a magnitude more sensitive than found in the current study under standard conditions. Hochmeister et al. (1) also conducted specificity, degradation, and contamination studies and determined that there was no effect on DNA recovery. This study was complimented by the work of Hermon et al. (2) who determined that sensitivity could be increased with a reduction of the buffer solution from 2 mL to 200 lL and also with an increase in incubation time. The maximum sensitivity that Hermon et al. (2) achieved was 1:1,000, even with an increase in incubation times, which they attribute to different volumes of blood used for the initial sampling. They also tested the specificity of the assay using different species of animals as well as other human body fluids.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.