Abstract
In their article, ‘‘AC frequency characteristics of coplanar impedance sensors as design parameters’’, Hong et al.1 present an analytical model for the calculation of the resistance between two electrodes in a microchannel. However, their measurements fit poorly with their suggested model, and they attribute the discrepancies to electrode thickness and a ‘fringing effect’. In this note we show that the discrepancies are due to the neglected channel height in their model, and that by including two additional transforma ions, the resistance between two
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have