Abstract

The purpose of this study was to investigate, using simulations, a method for improved contrast agent (CA) quantification in DCE-MRI. We developed a maximum likelihood estimator that combines the phase signal in the DCE-MRI image series with an additional CA estimate, e.g. the estimate obtained from magnitude data. A number of simulations were performed to investigate the ability of the estimator to reduce bias and noise in CA estimates. Noise levels ranging from that of a body coil to that of a dedicated head coil were investigated at both 1.5T and 3T. Using the proposed method, the root mean squared error in the bolus peak was reduced from 2.24 to 0.11 mM in the vessels and 0.16 to 0.08 mM in the tumor rim for a noise level equivalent of a 12-channel head coil at 3T. No improvements were seen for tissues with small CA uptake, such as white matter. Phase information reduces errors in the estimated CA concentrations. A larger phase response from higher field strengths or higher CA concentrations yielded better results. Issues such as background phase drift need to be addressed before this method can be applied in vivo.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.