Abstract

The value of the gas chromatography (GC) and selected ion flow tube mass spectrometry (SIFT-MS) combination for the analysis of trace gases is demonstrated by the quantification of acetone in air samples using the three precursor ions available to SIFT-MS, viz. H3O+, NO+ and O2+, and by the separation of the isomers 1-propanol and 2-propanol, and their analysis using H3O+ precursor ions. It is shown that the GC/SIFT-MS combination allows for accurate trace gas quantification obviating the regular, time-consuming calibrations that are usually required for the more commonly used detectors of GC systems, and the positive identification of isomers in mixtures that is often challenging using SIFT-MS alone. Thus, the GC/SIFT-MS combination paves the way to more confident analyses of complex mixtures such as exhaled breath.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.