Secondary electrospray corona discharge ionization (SECDI) combines the principles of secondary electrospray ionization (SESI) and corona discharge (CD) to achieve higher sensitivity, which is demonstrated through the detection of 2,4,6-trinitrotoluene (TNT) and 2,6-dinitrotoluene (2,6-DNT) vapors using ion mobility spectrometry (IMS). Using SECDI, enhancements in the IMS signal for TNT and 2,6-DNT vapors at trace concentrations are as much as 2–26 times that observed with CD or SESI alone. The enhancement in sensitivity is hypothesized to result from an increase in ionization efficiency driven by a higher number of reactant ions associated with SECDI compared to either technique individually. The ability of SECDI to achieve higher sensitivity without the aid of dopant molecules demonstrates its merit as an alternative ionization technique.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.