A transvenous implantable cardioverter defibrillator (ICD) was implanted in a patient with drug refractory ventricular fibrillation who had undergone latissimus dorsi cardiomyoplasty. The skeletal muscle was stimulated by a pulse train generator (cardiostimulator) implanted at the time of cardiomyoplasty. With proper programming of the devices neither adverse ICD-cardiostimulator interactions nor device malfunction were observed. Thus, the combined implantation of a cardiostimulator and an ICD is a feasible and safe therapeutic option.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.