Abstract

We report on a calibration-free dynamic carrier lifetime imaging technique yielding spatially resolved carrier lifetime maps of silicon wafers within data acquisition times of seconds. Our approach is based on infrared lifetime mapping (ILM), which exploits the proportionality between the measured infrared emission and the free carrier density. Dynamic ILM determines the lifetime analytically from the signal ratio of infrared camera images recorded directly after turning on an excitation source and after steady-state conditions are established within the sample. We investigate the applicability of dynamic infrared lifetime mapping on silicon wafers with rough surfaces, study the impact of injection dependencies, and examine the technical requirements for measuring low lifetime values in the range of microseconds. While the dynamic ILM approach is suitable for lifetimes exceeding 10μs, a combination with steady-state ILM is required to measure lifetime values in the range of 1μs. The injection dependence does not hamper a correct determination of the carrier lifetime by the dynamic evaluation procedure.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.