Abstract

The ultrasound Nakagami parameter map efficiently describes the statistical properties of the liver tissue. Using sliding windows in Nakagami imaging for monitoring radiofrequency ablation (RFA) is feasible. However, a limitation of conventional Nakagami imaging is the inability to visualize precise ablation (necrosis) zones by using a single sliding window. To accurately estimate ablation zones, we used window-modulated compounding (WMC) Nakagami imaging based on the synthesis of Nakagami images by using windows of different side lengths. Moreover, we proposed a new strategy for the postprocessing of Nakagami images by using Gaussian approximation (GAX) to approximate the cross-sectional areas of ablation zones. We heated 15 porcine liver samples for 12 min by using electrodes of tip lengths 0.5, 1.0, and 1.5 cm to ensure tissue necrosis after the ablation. The contour area of −6 dB in two-dimensional (2D) GAX images increased from 41 to 209 mm2, whereas that in the measures of section images increased from 40 to 265 mm2, with an increase in the electrode tips length from 0.5 to 1.5 cm. The coefficient of determination between the contour area of −6 dB in 2D GAX images and that in the measures of section images was 0.97, indicating a more satisfactory performance of GAX compared with that of polynomial approximation. The area estimated using GAX was similar to a circle, which conformed to the physical process of RFA.