Abstract

We characterize a 50-GHz comb generator with a sampling oscilloscope. With careful control of the input power, input harmonics, and comb generator temperature, we measure the output spectrum with a standard uncertainty of 0.1 dB and 0.5deg. We correct the measurements for time-base distortion, impedance mismatch, an inline attenuator, and the complex frequency response of the oscilloscope's sampler. We also report on the stability of the comb-generator spectrum and the effects of the recorded time windows, drive levels, and temperature. These results provide general guidelines for practioners of high-speed measurements and a benchmark for future inter-laboratory comparisons of harmonic phase-reference calibrations.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.