Abstract
Eleven cyanopropyl ("cyano") columns were characterized by means of a relationship developed originally for alkyl-silica columns. Compared to type-B alkyl-silica columns (i.e., made from pure silica), cyano columns are much less hydrophobic (smaller H), less sterically restricted (smaller S*), and have lower hydrogen-bond acidity (smaller A). Because sample retention is generally much weaker on cyano versus other columns (e.g., C8, C18), a change to a cyano column usually requires a significantly weaker mobile phase in order to maintain comparable values of k for both columns. For this reason, practical comparisons of selectivity between cyano and other columns (i.e., involving different mobile phases for each column) must take into account possible changes in separation due to the change in mobile phase, as well as change in the column.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.