Abstract

An in-vitro steady flow system was designed to determine the lowest flow velocities that can be detected by echographic colour flow imaging and spectral analysis. The flow detection level was determined hydrostatically by reducing the fluid level to below the point at which a flow signal was visible, then increasing the height until the colour flow reappeared. This was confirmed in all instances by spectral analysis. The height (angle-corrected velocity) of the spectral envelope was also determined. Mean volume flow was then obtained by using a graduated cylinder and a stop watch. The lowest hematocrit detectable was identified using a stepwise dilution of packed human red blood cells with 0.9% sodium chloride. Three different PVC-tubes with inner diameters of 2.1 mm, 1.1 mm and 0.51 mm were used.(ABSTRACT TRUNCATED AT 250 WORDS)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.