Abstract
Abstract Two proposed methods are reported for the quantitation of pirbuterol hydrochloride, namely, (i) colorimetric and (ii) titrimetric methods. The colorimetric method is based on coupling betweem diazotized sulphanilamide and pirbuterol hydrochloride. Under the optimum conditions studied, the coupling product exhibits a maximum at 440 nm. Linear relation between absorbance, A, and concentration of pirbuterol hydrochloride is in the range 5–40 μgml−1. The mean percentage obtained for capsules (ExirelR −15 mg) was 100.8 ± 0.7 whereas mean percentage recovery obtained for the authentic drug was 100.5 ± 0.8. The titrimetric procedure involves bromination of authentic pirbuterol in acid medium and residual titration of excess bromine. The stoichiometry of the reaction was investigated and infra-red analysis, of the bromoderivative was carried out. When applied to capsules the bromometric method gave mean percentage of 100.18 ± 2.25.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.