Abstract

Transthoracic color flow Doppler imaging has shown promise in evaluating Bjork Shiley mechanical prosthetic heart valves in a limited number of patients. 1 We present our color flow Doppler, pulsed and continuous wave Doppler findings in evaluating St. Jude Medical mitral and aortic valves.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.