Abstract

Purpose To assess the safety and efficacy of the Staar Collamer™ intraocular lens (IOL) 1 year postoperatively. Setting Fifteen private ophthalmology practices geographically distributed throughout the United States. Methods Six hundred eighty-six cases with a mean age of 72.1 years were enrolled in a 2-phase U.S. Food and Drug Administration (FDA) clinical study of the Collamer IOL implanted during phacoemulsification for cataract removal. The 12 month follow-up comprised data from 502 cases. Results Twelve months postoperatively, 96.2% of all cases and 95.6% without preexisting pathology achieved 20/40 or better best corrected visual acuity. Both results were better than FDA Grid values. No persistent sight-threatening complications, IOL dislocations, or IOL removals were reported. Conclusion Results show that the Collamer IOL is safe and effective for use in small incision cataract surgery.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.