Abstract

Abstract A rapid and precise method for the determination of fluoride in feeds employs HC1 extraction of the sample. Acetate buffer and sodium citrate are added to control pH and ionic strength. The amount of fluoride is calculated from a standard curve after measuring the potentials of standard and sample solutions. Eight collaborators participated in the study of the method. Statistical values on 3 pairs of samples were calculated. The standard deviation, precision, coefficient of variation, and bias, respectively, were: Pair 1, 0.005071, 0.001763, 3.09, 0.0034; Pair 2, 0.037122, 0.006475, 1.82, 0.0258; Pair 3, 0.034587, 0.013021, 2.63, 0.0227. The results from the proposed method agreed favorably with the values obtained by using the official final action AOAC method, 7.089. The average and standard deviation, respectively, for individual samples by the proposed method were: Sample 3, 0.049, 0.0029; Sample 4, 0.059, 0.0021; Sample 5, 0.334, 0.0114; Sample 6, 0.341, 0.0101; Sample 7, 0.511, 0.0219; Sample 8, 0.492, 0.0237. By the official method the values were: Sample 3, 0.049, 0.0041; Sample 4, 0.058, 0.0029; Sample 5, 0.334, 0.0055; Sample 6, 0.331, 0.0082; Sample 7, 0.517, 0.0183; Sample 8, 0.499, 0.0175. The ion selective electrode method has been adopted as official first action.