Abstract
A Flow Injection Analysis (FIA) procedure for the determination of trace amounts of mercury by the Cold Vapor Atomic Absorption (CVAA) method is presented. In order to make efficient use of the FIA technique, a new flow cell was developed to permit a rapid separation of elemental mercury from the carrier solution. The design of this device is based on the permeability of Hg 0 in commercially available Teflon (PTFE) tape, which acts as a membrane phase separator. The calibration curve is linear up to 70 ng.ml −1 ( A = 0.005 + 0.008 Hg(ng.ml −1); r = 0.9996) using a 15-cm open cell approach, with a detection limit of 1.4 ng. ml −1 (0.66 ng of mercury), considering a signal to noise ratio of three. The standard deviation is s = ±0.004 absorbance units, calculated at a level of 50 ng.ml −1 of Hg 2+. The sampling rate was 110 samples per h.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.