Abstract

Recently the US Food and Drug Administration has granted variances to select blood centers to supply cold-stored platelet components (CSP). In hemorrhage resuscitation warming of blood components with approved fluid warming devices is common. Pathogen-reduced apheresis platelet units were collected and stored in one of two ways: (1) CSP-I, (2) CSP-D. CSP-I were collected and immediately stored at 1-6°C until used. CSP-D were collected and stored at 20-24°C for 5 days and transferred to storage at 1-6°C until use. Aggregometry using arachidonic acid (AA), adenosine diphosphate (ADP) and collagen as agonists was performed on the unit samples before and after the units were infused through a Ranger blood-warming device. CSP-I, 23 units, had very high aggregation responses to all agonists (all ≥47.6 ± 20.7). There was a statistically significant reduction in ADP-induced aggregometry results from 55.1 ± 23.2 before compared to 33.5 ± 14.6 following infusion of the PLT through the blood warmer (p < .001). There were no differences in AA and collagen aggregometry results before and after the infusion of the platelets through the blood warmer. CSP-D had 5 of the 15 units with visible clotting in the bag. The 10 CSP-Ds studied had lower aggregation than all agonists before and after infusion through the blood-warming device (all ≤49.9 ± 35.9). We detected a statistically significant reduction in ADP-induced aggregometry in CSP-I run through a Ranger blood-warming device with no change with AA or collagen agonist aggregometry.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call