Abstract

Cold atmospheric pressure plasma (CAP) treatment has been shown to kill bacteria and remove bacterial biofilms from surfaces. Here we report the etch capacity of a linear discharge CAP device on Pseudomonas fluorescens biofilms. A 21 kHz, 1.4 kV RMS AC voltage applied to the CAP electrodes generated a hydrated Ar plasma between the plates, with the gas flow directing the plasma species toward the biological sample, causing both bacterial killing and etching of the biofilm. Typical discharge currents for a 2.4 cm long, 0.6 mm wide linear discharge device were 1-4.4 mA. Hydrated Ar flow gas was critical for removal of biofilm from a stainless steel substrate, while both hydrated and dry Ar + O2, Ar + air, O2 only, and air only flow gas mixtures did not cause etching at equivalent or greater discharge current intensities. A biofilm etch rate of > 2 μm/min was achieved, provided the plasma discharge was within 1-2 mm of the substrate surface and used a hydrated Ar gas flow of at least 5 LPM.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.