Abstract

Summary The objective of this study has been to examine the imaging properties of transmission electron microscopes when coherent non-planar illumination is used in conjunction with defocused specimens. This situation is reminiscent of what is commonly the case in electron microscopic phase-contrast studies of biologically relevant macromolecules, using a field emission gun as a coherent electron source. For the sake of simplicity, the imaging system has been idealized as a thin lens with properties that can be described by the Fresnel approximation of the Huygens-Fresnel principle. The resulting expressions show that the system magnification has a defocus dependent factor, as do the contrast transfer functions. These factors are normally not taken into account in conventional derivations. The defocus dependent factor can be minimized by using planar illumination. The factor approaches infinity as the crossover moves closer to the specimen, and it is in the region close to the specimen that this factor is most significant. These results can have serious implications for high-resolution single-particle cryo electron microscopy as this technique often relies on combining data taken at a range of defocus values.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.