Abstract

High-resolution imaging of trabecular bone aimed at analyzing the bone's microarchitecture is preferably performed with spin-echo-type pulse sequences. Unlike gradient echoes, spin-echoes are immune to artifactual broadening of trabeculae caused by local static field gradients near the bone-bone marrow interface and signal loss from chemical shift dephasing at k-space center. However, the previously practiced 3D fast large-angle spin-echo (FLASE) pulse sequence was found to be prone to a low-frequency modulation artifact in both the readout and slice direction. The artifact is caused by deviations in the effective flip angle of the nonselective 180 degrees pulse, which converts a fraction of the phase-encoded transverse magnetization to longitudinal magnetization. The latter recurs as transverse magnetization in the subsequent pulse sequence cycle forming a spurious stimulated echo. The objective of this work was to perform a k-space analysis of this steady-state artifact and propose two modifications of the original 3D FLASE that effectively remove it. The results of the simulations were in exact agreement with the experiments and the proposed remedy was found to eliminate the artifact.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.