Abstract

To investigate an anisotropic structural prior in morphology enabled dipole inversion (MEDI) for improving accuracy in quantitative susceptibility mapping (QSM). Anisotropic weighting (AW) was devised and implemented to incorporate orientation information into the edge agreement in the MEDI method. AW performance was compared with isotropic weighting by testing and validating on in vivo brain multiple orientation MRI data using COSMOS and the (33) component of the susceptibility tensor as reference. Suppressing streaking artifacts, AW improved not only QSM image quality but also accuracy in terms of RMSE (root mean square error), HFEN (high frequency error norm), SSIM (structural similarity index), and GDA (gradient direction agreement). In addition, it outperformed isotropic weighting in region of interest-based analysis. From a computational perspective, AW was as fast as isotropic weighting, taking approximately the same central processing unit times. Using AW in MEDI improves QSM accuracy compared with isotropic weighting. Magn Reson Med 79:1172-1180, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.