Abstract

A scheme of using phase-coded excitation and mismatched filter compression for high-frequency ultrasound imaging is presented in this paper. Biphase-coded pulses were constructed to excite the transducer. Received signals were compressed with mismatched filters optimized by minimizing peak-sidelobe-level (PSL). Both simulation and experiments were carried out to demonstrate the advantage of this technique. The simulation results demonstrated a possible sidelobe reduction (<−90 dB) with a slightly decrease of the signal-to-noise ratio of less than 1 dB compared with the compression using matched filters alone. The experimental results showed about 14 dB SNR improvement as well as −40 dB sidelobe level when the Barker-13 code excitation with 3-cycle sinusoidal wave carrier was used.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.