Abstract

A scheme of using phase-coded excitation and mismatched filter compression for high-frequency ultrasound imaging is presented in this paper. Biphase-coded pulses were constructed to excite the transducer. Received signals were compressed with mismatched filters optimized by minimizing peak-sidelobe-level (PSL). Both simulation and experiments were carried out to demonstrate the advantage of this technique. The simulation results demonstrated a possible sidelobe reduction (<−90 dB) with a slightly decrease of the signal-to-noise ratio of less than 1 dB compared with the compression using matched filters alone. The experimental results showed about 14 dB SNR improvement as well as −40 dB sidelobe level when the Barker-13 code excitation with 3-cycle sinusoidal wave carrier was used.