Abstract

ObjectivesTo review Cochlear's electrode portfolio and discuss the merits of current and future straight and perimodiolar electrode arrays. To present an update on implant reliability.MethodsPerformance and hearing preservation data from studies involving the Slim Straight (CI422), Hybrid L24 and Contour Advance electrode array were reviewed.ResultsWhile several studies in past found little difference in performance outcomes between subjects implanted with perimodiolar and straight arrays, recent studies demonstrated that proximity to the modiolus is correlated with better performance. Hearing threshold increase was lowest with the Hybrid L24, closely followed by the slim straight array and was largest with the Contour Advance array. The CI24RE receiver-stimulator used for the three arrays had a cumulative survival of 99% at eight years post implantation.ConclusionsCombining the hearing preservation benefits of slim straight arrays with perimodiolar proximity is the design objective of Cochlear's next generation electrodes.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.